TY-DCL400 Residual Chlorine/Chlorine Dioxide Online Analyzer
Contact Info
- Add:西安市高新区唐延南路11号, Zip:
- Contact: 谭东东
- Tel:13720774318
- Email:509857967@qq.com
Other Products
Product Introduction:
The residual chlorine/clo2 sensor adopts an advanced constant voltage principle to measure residual chlorine/clo2 in water. This method applies a stable potential between the polarized electrode and the reference electrode, where different measured components generate varying current intensities under this potential. The instrument calculates the concentration of the measured component by collecting and analyzing the current signal.
The residual chlorine/clo2 electrode has a simple structure, is easy to clean and replace, and does not require membrane or reagent replacement during use, ensuring simple maintenance and long-term stability, reliability, and accuracy of the instrument.
Typical Applications:
lUsed for residual chlorine/clo2 measurement in water treatment plants, drinking water distribution networks, and swimming pools.
Sensor Technical Parameters:
Principle | Constant Voltage Method |
Measurement Range | 0-2mg/L, 0-20 mg/L (optional) |
Calibration | Two-point calibration |
Slope | 78%--155% |
Temperature Compensation | Manual or automatic temperature compensation |
Temperature Coefficient | 2%/℃ |
Sample pH Range | 5-8 |
Flow Rate Requirement | 30-60L/h |
Operating Temperature | 0 to 50℃ (non-freezing) |
Electrode Dimensions | Diameter 13.5mm * Length 178.5mm |
Installation Thread Size | PG13.5 |
Electrode Material | 316 stainless steel housing |
Cable Length | Standard 5-meter cable, extendable up to 20 meters |
DS-115T Controller Technical Parameters:
Display | 128*64 graphic dot matrix LCD with strong LED backlight, operable under direct sunlight |
Power Supply | AC power supply: 85-500VAC (50/60HZ); DC power supply: 9~36VDC |
Output | 3 channels of 4-20mA |
Relay | Three relays, programmatically set response parameters and values |
Communication Protocol | MODBUS RS485 |
Main Material | PA66+GF25+FR (upper cover); powder-coated aluminum alloy (lower shell) |
Storage Temperature | -20 to 70℃ |
Operating Temperature | -15 to 60℃ |
Protection Rating | IP65/ NEMA4X |
Dimensions | 145*125*162mm(L*W*H) |
Weight | 1.35KG |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 陕西天韵 |
| Spec: | TY-DCL400 |
| Stock: | |
| Origin: | China / Shaanxi / Xianshi |